Endo Gator for Pump Irrigation Units Endo Gator for Endostat II & Olympus OFP Reusable adapter for Olympus 140/160/180 series scopes Endo SmartCap 140/160/180 series scopes Endo SmartCap for Olympus Endo SmartCap for Pentax Endo SmartCap for Fujinon Sterile water bottle holder


Product #100160

Also available: 100160 CO2

  1. Open the sterile peel pack and remove the Endo SmartCap. Open a bottle of sterile water (500 cc or one liter). The Endo SmartCap will fit Abbott, Baxter and MacGaw water bottles.

  2. Drop the metal weighted tube into the sterile water bottle and firmly tighten the Endo SmartCap to ensure a tight seal.
     
  3. Plug the distal tip of the Endo SmartCap into the scope by twisting while pushing firmly. Reconnect the air inlet tube into the processor with a firm twist until the connector clicks into place. Check the air/water flow. If the water pressure is low, check to ensure the Endo SmartCap has a tight seal.
     
  4. When the procedure is finished, turn off the air pressure on the scope processor then clamp off the tubing with the pinch clip before disconnecting the Endo SmartCap.